ABSTRACT: Purpose
Multichannel transcranial magnetic stimulation (TMS)1 is an emerging technology that allows multiple sites to be stimulated simultaneously or sequentially under electronic control without movement of the coils. A multichannel TMS/MRI head coil array for 3 Tesla is currently under development to mitigate challenges of concurrent TMS/fMRI as well as enable potential new applications. The influence of the multichannel TMS system on the MR image quality and safety must be carefully investigated.Methods
A standard birdcage volume coil for 3 Tesla systems was simulated using a commercial numerical electromagnetic solver. Two setups, consisting of 1) a MR-compatible TMS coil, and 2) a 3-axis TMS coil array, were simulated to quantify changes in the transmit field B1+ and the SAR. A realistically shaped homogeneous head model was used in the computations.Results
The stimulation coils produced enhancements and attenuations on the transmit field with effects greater than 5% up to 2.4 cm and 3.3 cm under the scalp for the MR-compatible TMS coil and 3-axis TMS coil array, respectively. The 10 g-SAR distribution did not change significantly in either of the cases; however, the nominal SAR maximum locus was shifted between existing hot spots.Conclusion
The simulated B1+ variations found near the TMS coils indicate the possibility of inducing sequence-dependent image artefacts predominatly limited to the vicinity of the coil(s). However, we conclude that neither the MR-compatible commercial TMS coil nor the 3-axis TMS coil array siginificantly elevate SAR in the head or neck beyond accepted safety limits.
SUBMITTER: Navarro de Lara LI
PROVIDER: S-EPMC7180091 | biostudies-literature |
REPOSITORIES: biostudies-literature