In-House Standard Method for Molecular Characterization of Dissolved Organic Matter by FT-ICR Mass Spectrometry.


ABSTRACT: Electrospray ionization (ESI) coupled with Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS) has been widely used for molecular characterization of dissolved organic matter (DOM). However, ESI FT-ICR MS generally has poor repeatability and reproducibility because of its inherent ionization mechanism and structural characteristics, which severely hindered its application in quantitative analysis of complex mixtures. In this article, we developed an in-house standard method for molecular characterization of DOM by ESI FT-ICR MS. Instead of obtaining reproducible results by determining the instrument parameters, we adopted an approach of object control on the mass spectrum to solve the problem of poor reproducibility. The mass peak shape, resolution, and relative intensity distribution of a natural organic matter standard were adjusted by optimizing the operating conditions to obtain a repeatable result. The quality control sample was run 26 times by the different operators in a 6-month-long period to evaluate the reproducibility. Results showed that the relative standard deviation (%) of repeatability and reproducibility are 1.02 and 2.35 for average H/C, respectively. The in-house standard method has been validated and successfully used for the characterization of more than 4000 DOM samples, which is transferable to other laboratories.

SUBMITTER: He C 

PROVIDER: S-EPMC7254807 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

In-House Standard Method for Molecular Characterization of Dissolved Organic Matter by FT-ICR Mass Spectrometry.

He Chen C   Zhang Yahe Y   Li Yunyun Y   Zhuo Xiaocun X   Li Yuguo Y   Zhang Chuanlun C   Shi Quan Q  

ACS omega 20200514 20


Electrospray ionization (ESI) coupled with Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR MS) has been widely used for molecular characterization of dissolved organic matter (DOM). However, ESI FT-ICR MS generally has poor repeatability and reproducibility because of its inherent ionization mechanism and structural characteristics, which severely hindered its application in quantitative analysis of complex mixtures. In this article, we developed an in-house standard method f  ...[more]

Similar Datasets

| S-EPMC6617137 | biostudies-literature
| S-EPMC9260710 | biostudies-literature
| S-EPMC7709006 | biostudies-literature
| S-EPMC3739293 | biostudies-literature
| S-EPMC8073082 | biostudies-literature
| S-EPMC7081437 | biostudies-literature
| S-EPMC8892831 | biostudies-literature
| S-EPMC4095805 | biostudies-literature
| S-EPMC7031845 | biostudies-literature
| S-EPMC5353847 | biostudies-literature