Comparison of internal target volumes defined by three-dimensional, four-dimensional, and cone-beam computed tomography images of a motion phantom.


ABSTRACT:

Background

To explore the variations of the gross tumor volume (GTV) from different three-dimensional computed tomography (3DCT) scan modes and the consistency of internal target volume (ITV) between different 3DCT, cone-beam computed tomography (CBCT) and four-dimensional computed tomography (4DCT) scan, a study using a motion phantom simulating sinusoidal movement was conducted.

Methods

For three 3DCT scan modes: the GTV was contoured, and ITVI was generated on the basis of GTV with a 10-mm margin while ITVII and ITVIII with a 0-mm margin on the motion direction. ITVCBCT and ITVMIP were contoured on the images of CBCT and maximum-intensity projection (MIP) reconstructed 4DCT images. The centroid position shifts of ITVs were analyzed. The volume consistency between ITVI, ITVII, ITVIII and ITVMIP were evaluated by calculating the Dice similarity coefficient (Dsc) and the value of Δ Volume (ΔV). Furthermore, the 3DCT and CBCT images from 12 NSCLC patients were retrospectively collected, then the Dsc and ΔV were calculated.

Results

The mean ± standard deviation of centroid position of ITVI, ITVII and ITVIII were 2.3±4.7, 2.6±4.0, and 1.0±1.4 mm, respectively. The mean ± standard deviation of Dsc between ITVI, ITVII, ITVIII and ITVMIP were 0.78±0.77, 0.86±0.1, and 0.94±0.05, respectively. The ΔV of ITVI, ITVII, ITVIII were 29.67%, 17.22%, and 6.46%, respectively. The ITV from CBCT showed a deduction rate of 3.1-9.3% compared to 4DCT. For the patients, the mean Dsc andΔV between ITVI and ITVCBCT were 0.50 and 60.76%.

Conclusions

The GTV acquired from 3DCT scan mode I possessed great deviation of centroid position and target volume. ITV on the basis of this GTV was significantly larger than ITVMIP. A good similarity was showed between ITVIII and ITVMIP, 4DCT is still a golden standard for the ITV delineation, but in the absence of 4DCT, image from 3DCT scan mode III and KV-CBCT may be considered for ITV delineation with caution.

SUBMITTER: Wang Y 

PROVIDER: S-EPMC7729323 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of internal target volumes defined by three-dimensional, four-dimensional, and cone-beam computed tomography images of a motion phantom.

Wang Yu Y   Liu Tao T   Chen Huiqin H   Bai Penggang P   Zhan Qiong Q   Liang Xiaohua X  

Annals of translational medicine 20201101 22


<h4>Background</h4>To explore the variations of the gross tumor volume (GTV) from different three-dimensional computed tomography (3DCT) scan modes and the consistency of internal target volume (ITV) between different 3DCT, cone-beam computed tomography (CBCT) and four-dimensional computed tomography (4DCT) scan, a study using a motion phantom simulating sinusoidal movement was conducted.<h4>Methods</h4>For three 3DCT scan modes: the GTV was contoured, and ITV<sub>I</sub> was generated on the ba  ...[more]

Similar Datasets

| S-EPMC5119621 | biostudies-literature
| S-EPMC10278033 | biostudies-literature
| S-EPMC7125134 | biostudies-literature
| S-EPMC9958024 | biostudies-literature
| S-EPMC4032412 | biostudies-other
| S-EPMC7807672 | biostudies-literature
| S-EPMC7807636 | biostudies-literature
| S-EPMC7194353 | biostudies-literature
| S-EPMC7500863 | biostudies-literature
| S-EPMC6400385 | biostudies-literature