Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer.


ABSTRACT:

Background

In recurrent differentiated thyroid cancer patients, detectability in 124I PET is limited for lesions with low radioiodine uptake. We assess the improvements in lesion detectability and image quality between three generations of PET scanners with different detector technologies. The results are used to suggest an optimized protocol.

Methods

Datasets of 10 patients with low increasing thyroglobulin or thyroglobulin antibody levels after total thyroidectomy and radioiodine therapies were included. PET data were acquired and reconstructed on a Biograph mCT PET/CT (whole-body, 4-min acquisition time per bed position; OSEM, OSEM-TOF, OSEM-TOF+PSF), a non-TOF Biograph mMR PET/MR (neck region, 4 min and 20 min; OSEM), and a new generation Biograph Vision PET/CT (whole-body, 4 min; OSEM, OSEM-TOF, OSEM-TOF+PSF). The 20-min image on the mMR was used as reference to calculate the detection efficacy in the neck region. Image quality was rated on a 5-point scale.

Results

All detected lesions were in the neck region. Detection efficacy was 8/9 (Vision OSEM-TOF and OSEM-TOF+PSF), 4/9 (Vision OSEM), 3/9 (mMR OSEM and mCT OSEM-TOF+PSF), and 2/9 (mCT OSEM and OSEM-TOF). Median image quality was 4 (Vision OSEM-TOF and OSEM-TOF+PSF), 3 (Vision OSEM, mCT OSEM-TOF+PSF, and mMR OSEM 20-min), 2 (mCT OSEM-TOF), 1.5 (mCT OSEM), and 1 (mMR OSEM 4 min).

Conclusion

At a clinical standard acquisition time of 4 min per bed position, the new generation Biograph Vision using a TOF-based image reconstruction demonstrated the highest detectability and image quality and should, if available, be preferably used for imaging of low-uptake lesions. A prolonged acquisition time for the mostly affected neck region can be useful.

SUBMITTER: Kersting D 

PROVIDER: S-EPMC7884562 | biostudies-literature | 2021 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparing lesion detection efficacy and image quality across different PET system generations to optimize the iodine-124 PET protocol for recurrent thyroid cancer.

Kersting David D   Jentzen Walter W   Sraieb Miriam M   Costa Pedro Fragoso PF   Conti Maurizio M   Umutlu Lale L   Antoch Gerald G   Nader Michael M   Herrmann Ken K   Fendler Wolfgang Peter WP   Rischpler Christoph C   Weber Manuel M  

EJNMMI physics 20210215 1


<h4>Background</h4>In recurrent differentiated thyroid cancer patients, detectability in <sup>124</sup>I PET is limited for lesions with low radioiodine uptake. We assess the improvements in lesion detectability and image quality between three generations of PET scanners with different detector technologies. The results are used to suggest an optimized protocol.<h4>Methods</h4>Datasets of 10 patients with low increasing thyroglobulin or thyroglobulin antibody levels after total thyroidectomy and  ...[more]

Similar Datasets

| S-EPMC8410931 | biostudies-literature
| S-EPMC3356447 | biostudies-literature
| S-EPMC7876195 | biostudies-literature
| S-EPMC8003652 | biostudies-literature
| S-EPMC3868598 | biostudies-literature
| S-EPMC5015757 | biostudies-literature
| S-EPMC5671737 | biostudies-literature
| S-EPMC4768638 | biostudies-literature
| S-EPMC4779400 | biostudies-literature
| S-EPMC5493009 | biostudies-literature