Unknown

Dataset Information

0

Proof of Concept: Measuring Aortic Annulus Resistance by Means of Pressure-Volume Curves During Balloon Inflation to Guide Transcatheter Aortic Valve Implantation.


ABSTRACT: This study assessed the basic working principle to measure aortic annulus resistance during balloon inflation for transcatheter aortic valve implantation (TAVI), by acquisition of pressure-volume curve for a guided semi-automatic implantation. A modular bench-system was used which allows the incremental inflation of valvuloplasty balloons by means of a stepper-motor driven linear axis with simultaneous recording of the pressure changes inside the system. Different porcine aortic xenografts were assessed by use of a non-compliant valvuloplasty balloon. In a second step transcatheter aortic stents were implanted inside target sized xenografts. The recorded pressure volume-curves showed that the system can accurately differentiate between different xenografts and assess the quality of the tissue rendering real-time analysis of pressure-volume curves during balloon-inflation possible, which has the potential to optimize the implantation procedure by direct adaptation to the patient specific anatomy and characteristics. Further investigations and development are warranted.

SUBMITTER: Noterdaeme T 

PROVIDER: S-EPMC8119645 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC5889984 | biostudies-literature
| S-EPMC6242918 | biostudies-literature
| S-EPMC5590871 | biostudies-literature
| S-EPMC10562087 | biostudies-literature
| S-EPMC8639544 | biostudies-literature
| S-EPMC9700572 | biostudies-literature
| S-EPMC5210340 | biostudies-literature
| S-EPMC7925971 | biostudies-literature
| S-EPMC7428568 | biostudies-literature
| S-EPMC7977026 | biostudies-literature