ABSTRACT: Background
Up to 7.5% of tined-lead removals in patients having sacral neuromodulation (SNM) therapy are associated with a lead breakage. It is still unclear what adverse effects can be caused by unretrieved fragments. The aim of our study was to describe the lead removal technique we have been using for the last 2 years in our centre.Methods
We retrospectively enrolled patients who had lead removal between January 2018 and January 2020 using our standardized technique. The novelty of the technique is in the use of the straight stylet, which is available in the quadripolar tined-lead kit. The stylet gives the electrode greater stiffness, reducing interactions with surrounding tissues and probability of damage or breakage during removal.Results
In 59 patients (42 women, mean age 57.2 years [range 40-79 years]) the lead was removed using our standardized technique. In 44 of 59 patients, the tined-lead was removed within 2 months from the SNM-test, due to lack of beneficial effects. In 15 patients the electrode was removed because of failure of definitive implantation. Meantime from definitive implantable pulse generator (IPG) implantation to lead removal was 67.9 months. We recorded only 1 case of lead-breakage during removal: a female patient with a non-tined lead fixed on sacral bone, placed 18 years previously using an open technique.Conclusions
Lead breakage during removal is not uncommon and adverse effects of retained fragments may occur. Our technique has been safely used for the last 2 years in our centre, with no episodes of lead breakage or retained fragments, except for one non-tined electrode.
SUBMITTER: Agnello M
PROVIDER: S-EPMC8289802 | biostudies-literature |
REPOSITORIES: biostudies-literature