Lead related complications in quadripolar versus bipolar left ventricular leads.
Ontology highlight
ABSTRACT: Quadripolar left ventricular (LV) leads are capable of pacing from four different electrodes which allows for easier and more stable intra-operative lead positioning with optimal pacing parameters. We therefore investigated the rate of combined intra-operative and post-operative LV lead related events in quadripolar vs. bipolar LV lead cardiac resynchronization therapy (CRT) recipients in the real world setting.We retrospectively collected data for N = 1441 patients at our institution implanted with quadripolar (n = 292) or bipolar (n = 1149) LV leads from 2012 to 2014 and followed them to the primary end-point of composite lead outcome defined as intra-operative lead implant failure or post-operative lead dislodgement or deactivations.Patients implanted with a quadripolar lead were younger (70.6 ± 11.4 vs 72.5 ± 11.6, p = 0.014) and had higher incidence of diabetes (41.8% vs 32.8%, p = 0.004) compared to those with bipolar leads. All other baseline characteristics were comparable. Patients implanted with a quadripolar were significantly less likely to reach the primary endpoint in the first 12 months after LV lead implantation (Hazard Ratio 0.22, 95% Confidence Interval 0.08-0.60, p = 0.001). There were no differences between the two groups in rates of hospitalization for any cause or in mortality.In this real world study, quadripolar LV leads have significantly lower rates of implantation failure and post-operative lead dislodgement or deactivation. These results have important clinical implications to CRT recipients.
SUBMITTER: Rijal S
PROVIDER: S-EPMC5357860 | biostudies-other | 2017 Jan - Feb
REPOSITORIES: biostudies-other
ACCESS DATA