Unknown

Dataset Information

Specialized computational methods for denoising, B1 correction, and kinetic modeling in hyperpolarized 13 C MR EPSI studies of liver tumors.


ABSTRACT:

Purpose

To develop a novel post-processing pipeline for hyperpolarized (HP) 13 C MRSI that integrates tensor denoising and B1+ correction to measure pyruvate-to-lactate conversion rates (kPL ) in patients with liver tumors.

Methods

Seven HP 13 C MR scans of progressing liver tumors were acquired using a custom 13 C surface transmit/receive coil and the echo-planar spectroscopic imaging (EPSI) data analysis included B0 correction, tensor rank truncation, and zero- and first-order phase corrections to recover metabolite signals that would otherwise be obscured by spectral noise as well as a correction for inhomogeneous transmit ( B1+ ) using a B1+ map aligned to the coil position for each patient scan. Processed HP data an

SUBMITTER: Lee PM 

PROVIDER: S-EPMC8565779 | biostudies-literature | 2021 Nov

REPOSITORIES: biostudies-literature