Unknown

Dataset Information

0

FPGA Integrated Optofluidic Biosensor for Real-Time Single Biomarker Analysis.


ABSTRACT: Integrated optofluidic biosensors can fill the need for sensitive, amplification-free, multiplex single molecule detection which is relevant for containing the spread of infectious diseases such as COVID-19. Here, we demonstrate a rapid sample-to-answer scheme that uses a field programmable gate array (FPGA) to enable live monitoring of single particle fluorescence analysis on an optofluidic chip. Fluorescent nanobeads flowing through a micro channel are detected with 99% accuracy and particle concentrations in clinically relevant ranges from 3.4×104 to 3.4 × 106/ml are determined within seconds to a few minutes without the need for post-experiment data extraction and analysis. In addition, other extract salient experimental parameters such as dynamic flow rate changes can be monitored in real time. The sensor is validated with real-time fluorescence detection of single bacterial plasmid DNA at attomolar concentrations, showing excellent promise for implementation as a point of care (POC) diagnostic tool.

SUBMITTER: Sampad MJN 

PROVIDER: S-EPMC8658630 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC4629162 | biostudies-literature
| S-EPMC4933911 | biostudies-literature
| S-EPMC9420106 | biostudies-literature
| S-EPMC3075913 | biostudies-literature
| S-EPMC3622590 | biostudies-literature
| S-EPMC10311125 | biostudies-literature
| S-EPMC7584764 | biostudies-literature
2018-12-31 | GSE122233 | GEO
| S-EPMC8374455 | biostudies-literature
| S-EPMC5537690 | biostudies-literature