Reduced field-of-view and multi-shot DWI acquisition techniques: Prospective evaluation of image quality and distortion reduction in prostate cancer imaging.


ABSTRACT:

Objectives

To prospectively compare image quality and apparent diffusion coefficient (ADC) quantification for reduced field-of-view (rFOV)- and multi-shot echo-planar imaging (msEPI)-based diffusion weighted imaging (DWI), using single-shot echo-planar-imaging (ssEPI) DWI as the reference.

Methods

Under IRB approval and after informed consent, msEPI, rFOV, and ssEPI DWI acquisitions were prospectively added to clinical prostate MRI exams at 3.0 T. Image distortion was quantitatively evaluated by root-mean-squared displacement (dr.m.s.). Histogram-based quantitative ADC parameters were compared in a sub-set of patients for proven sites of prostate cancer and matched non-cancerous prostate. Three radiologists also independently evaluated the DWI sequences for subjective image quality and distortion/artifact on a 5-point Likert scale.

Results

Twenty-five patients were included (15 with proven sites of cancer). Average dr.m.s. demonstrated a small but statistically significant reduction in distortion for both rFOV and msEPI relative to ssEPI. Quantitative ADC parameters for prostate tumors demonstrated no significant difference across the 3 DWI acquisitions and each acquisition demonstrated a statistically significant decrease in mean ADC for tumor compared to normal prostate. Qualitative reader assessment demonstrated favorable image quality for rFOV and msEPI, more notable for msEPI.

Conclusions

rFOV and msEPI DWI techniques achieved reduction in image distortion, improvement in image quality, and maintained reproducible ADC quantification compared to the standard ssEPI.

SUBMITTER: Lawrence EM 

PROVIDER: S-EPMC9523455 | biostudies-literature | 2022 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Reduced field-of-view and multi-shot DWI acquisition techniques: Prospective evaluation of image quality and distortion reduction in prostate cancer imaging.

Lawrence Edward M EM   Zhang Yuxin Y   Starekova Jitka J   Wang Zihan Z   Pirasteh Ali A   Wells Shane A SA   Hernando Diego D  

Magnetic resonance imaging 20220806


<h4>Objectives</h4>To prospectively compare image quality and apparent diffusion coefficient (ADC) quantification for reduced field-of-view (rFOV)- and multi-shot echo-planar imaging (msEPI)-based diffusion weighted imaging (DWI), using single-shot echo-planar-imaging (ssEPI) DWI as the reference.<h4>Methods</h4>Under IRB approval and after informed consent, msEPI, rFOV, and ssEPI DWI acquisitions were prospectively added to clinical prostate MRI exams at 3.0 T. Image distortion was quantitative  ...[more]

Similar Datasets

| S-EPMC5607067 | biostudies-literature
| S-EPMC9880109 | biostudies-literature
| S-EPMC7392767 | biostudies-literature
| S-EPMC7965571 | biostudies-literature
| S-EPMC7808647 | biostudies-literature
| S-EPMC3431233 | biostudies-literature
| S-EPMC8338849 | biostudies-literature
| S-EPMC4756687 | biostudies-literature
| S-EPMC4090294 | biostudies-literature
| S-EPMC9343635 | biostudies-literature