Real-time detection of airborne viruses on a mass-sensitive device.
Ontology highlight
ABSTRACT: We present real-time detection of airborne Vaccinia viruses using quartz crystal microbalance (QCM) in an integrated manner. Vaccinia viruses were aerosolized and neutralized using an electrospray aerosol generator, transported into the QCM chamber, and captured by a QCM crystal. The capture of the viruses on the QCM crystal resulted in frequency shifts proportional to the number of viruses. The capture rate varied linearly with the concentration of initial virus suspensions (8.5x10(8)-8.5x10(10) particlesml) at flow rates of 2.0 and 1.1 lmin. This work demonstrates the general potential of mass sensitive detection of nanoscale biological entities in air.
SUBMITTER: Lee J
PROVIDER: S-EPMC2682748 | biostudies-other | 2008 Jul
REPOSITORIES: biostudies-other
ACCESS DATA