Scanning mass spectrometry probe: a scanning probe electrospray ion source for imaging mass spectrometry of submerged interfaces and transient events in solution.


ABSTRACT: The scanning mass spectrometry (SMS) probe is a new electrospray ion source. Motivated by the need for untargeted chemical imaging of dynamic events in solution, we have exploited an approach to electrospray ionization (ESI) that allows continuous sampling from a highly localized volume (approximately picoliters) in a liquid environment, softly ionizes molecules in the sample to render them amenable for mass spectrometric analysis, and sends the ions to the mass spectrometer. The key underlying concepts for our approach are (1) treating the electrospray capillary inlet as a chemical scanning probe and (2) locating the electrospray point as close as possible to the sampling point, thus providing the shortest response time possible. This approach enables chemical monitoring or imaging of submerged interfaces, providing access to details of spatial heterogeneity and temporal changes within liquid samples. It also permits direct access to liquid/ liquid interfaces for ESI-MS analysis. In this letter we report the first demonstrations of these capabilities of the SMS probe and describe some of the probe's basic characteristics.

SUBMITTER: Kottke PA 

PROVIDER: S-EPMC2800854 | biostudies-literature | 2010 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Scanning mass spectrometry probe: a scanning probe electrospray ion source for imaging mass spectrometry of submerged interfaces and transient events in solution.

Kottke Peter A PA   Degertekin F Levent FL   Fedorov Andrei G AG  

Analytical chemistry 20100101 1


The scanning mass spectrometry (SMS) probe is a new electrospray ion source. Motivated by the need for untargeted chemical imaging of dynamic events in solution, we have exploited an approach to electrospray ionization (ESI) that allows continuous sampling from a highly localized volume (approximately picoliters) in a liquid environment, softly ionizes molecules in the sample to render them amenable for mass spectrometric analysis, and sends the ions to the mass spectrometer. The key underlying  ...[more]

Similar Datasets

| S-EPMC6310154 | biostudies-literature
| S-EPMC4055847 | biostudies-literature
| S-EPMC2631554 | biostudies-literature
| S-EPMC7216899 | biostudies-literature
| S-EPMC5157838 | biostudies-literature
| S-EPMC7013221 | biostudies-literature
| S-EPMC3704159 | biostudies-literature
| S-EPMC3176668 | biostudies-literature
| S-EPMC6422691 | biostudies-literature
| S-EPMC7108174 | biostudies-literature