Unknown

Dataset Information

0

Precise beam-tilt alignment and collimation are required to minimize the phase error associated with coma in high-resolution cryo-EM.


ABSTRACT: Electron microscopy at a resolution of 0.4nm or better requires more careful adjustment of the illumination than is the case at a resolution of 0.8nm. The use of current-axis alignment is not always sufficient, for example, to avoid the introduction of large phase errors, at higher resolution, due to axial coma. In addition, one must also ensure that off-axis coma does not corrupt the data quality at the higher resolution. We particularly emphasize that the standard CTF correction does not account for the phase error associated with coma. We explain the cause of both axial coma and the typically most troublesome component of off-axis coma in terms of the well-known shift of the electron diffraction pattern relative to the optical axis that occurs when the illumination is not parallel to the axis. We review the experimental conditions under which coma causes unacceptably large phase errors, and we discuss steps that can be taken when setting up the conditions of illumination, so as to ensure that neither axial nor off-axis coma is a problem.

SUBMITTER: Glaeser RM 

PROVIDER: S-EPMC3056894 | biostudies-literature | 2011 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Precise beam-tilt alignment and collimation are required to minimize the phase error associated with coma in high-resolution cryo-EM.

Glaeser Robert M RM   Typke Dieter D   Tiemeijer Peter C PC   Pulokas James J   Cheng Anchi A  

Journal of structural biology 20101221 1


Electron microscopy at a resolution of 0.4nm or better requires more careful adjustment of the illumination than is the case at a resolution of 0.8nm. The use of current-axis alignment is not always sufficient, for example, to avoid the introduction of large phase errors, at higher resolution, due to axial coma. In addition, one must also ensure that off-axis coma does not corrupt the data quality at the higher resolution. We particularly emphasize that the standard CTF correction does not accou  ...[more]

Similar Datasets

| S-EPMC5949096 | biostudies-literature
| S-EPMC8098478 | biostudies-literature
| S-EPMC7924229 | biostudies-literature
| S-EPMC9733953 | biostudies-literature
| S-EPMC4585738 | biostudies-literature
| S-EPMC7642776 | biostudies-literature
| S-EPMC8027200 | biostudies-literature
| S-EPMC7768659 | biostudies-literature
| S-EPMC3684049 | biostudies-literature
| S-EPMC7613808 | biostudies-literature