Accelerated 3D T2 w-imaging of the prostate with 1-millimeter isotropic resolution in less than 3 minutes.


ABSTRACT:

Purpose

To achieve 3D T2 w imaging of the prostate with 1-mm isotropic resolution in less than 3 min.

Methods

We devised and implemented a 3D T2 -prepared multishot balanced steady state free precession (T2 prep-bSSFP) acquisition sequence with a variable density undersampled trajectory combined with a total variation regularized iterative SENSE (TV-SENSE) reconstruction. Prospectively undersampled images of the prostate (acceleration factor R = 3) were acquired in 11 healthy subjects in an institutional review board-approved study. Image quality metrics (subjective signal-to-noise ratio, contrast, sharpness, and overall prostate image quality) were evaluated by 2 radiologists. Scores of the proposed accelerated sequence were compared using the Wilcoxon signed-rank and Kruskal-Wallis non-parametric tests to prostate images acquired using a fully sampled 3D T2 prep-bSSFP acquisition, and with clinical standard 2D and 3D turbo spin echo (TSE) T2 w acquisitions. A P-value < 0.05 was considered significant.

Results

The 3× accelerated 3D T2 prep-bSSFP images required a scan time (min:s) of 2:45, while the fully sampled 3D T2 prep-bSSFP and clinical standard 3D TSE images were acquired in 8:23 and 7:29, respectively. Image quality scores (contrast, sharpness, and overall prostate image quality) of the accelerated 3D T2 prep-bSSFP, fully sampled T2 prep-bSSFP, and clinical standard 3D TSE acquisitions along all 3 spatial dimensions were not significantly different (P > 0.05).

Conclusion

3D T2 w images of the prostate with 1-mm isotropic resolution can be acquired in less than 3 min, with image quality that is comparable to a clinical standard 3D TSE sequence but only takes a third of the acquisition time.

SUBMITTER: Vidya Shankar R 

PROVIDER: S-EPMC6563534 | biostudies-literature | 2019 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accelerated 3D T<sub>2</sub> w-imaging of the prostate with 1-millimeter isotropic resolution in less than 3 minutes.

Vidya Shankar Rohini R   Roccia Elisa E   Cruz Gastao G   Neji Radhouene R   Botnar René R   Prezzi Davide D   Goh Vicky V   Prieto Claudia C   Dregely Isabel I  

Magnetic resonance in medicine 20190421 2


<h4>Purpose</h4>To achieve 3D T<sub>2</sub> w imaging of the prostate with 1-mm isotropic resolution in less than 3 min.<h4>Methods</h4>We devised and implemented a 3D T<sub>2</sub> -prepared multishot balanced steady state free precession (T<sub>2</sub> prep-bSSFP) acquisition sequence with a variable density undersampled trajectory combined with a total variation regularized iterative SENSE (TV-SENSE) reconstruction. Prospectively undersampled images of the prostate (acceleration factor R = 3)  ...[more]

Similar Datasets

| S-EPMC6899588 | biostudies-literature
| S-EPMC7161114 | biostudies-literature
| S-EPMC6617822 | biostudies-literature
| S-EPMC6107445 | biostudies-literature
| S-EPMC7782498 | biostudies-literature
| S-EPMC8032210 | biostudies-literature
| S-EPMC4224117 | biostudies-literature
| S-EPMC6851769 | biostudies-literature
| S-EPMC8277490 | biostudies-literature
| S-EPMC3426855 | biostudies-other