Unknown

Dataset Information

0

Second window near-infrared dosimeter (NIR2D) system for radiation dosimetry.


ABSTRACT: Fiber-coupled scintillation dosimeters are a cost-effective alternative to the conventional ion chambers in radiation dosimetry. However, stem effects from optical fibers such as Cerenkov radiation incur significant errors in the readout signal. Here we introduce a second near-infrared window dosimeter, dubbed as NIR2D, that can potentially be used as real-time radiation detector for clinical megavoltage beams. Lanthanide-based rare-earth NaYF4 nano-phosphors doped with both erbium and cerium elements were synthesized, and a compact 3D printed reader device integrated with a photodetector and data acquisition system was designed. The performance of the NIR2D was tested using a pre-clinical orthovoltage radiation source and a clinical megavoltage radiation source. The system was tested for dose linearity (100, 200, 600 MU), dose rate dependency (100, 200, 400, 600 MU min-1), and energy dependency (6, 10, 15 MV). Test results with the clinical linear accelerator demonstrated excellent dose linearity and dose rate independency when exposed to 6 MV linac beams-both data follows a linear trendline with R2 > 0.99. On the other hand, the NIR2D was energy dependent, where the readout dropped by 9% between 6 and 15 MV. For stem effects, we observed a finite Cerenkov contribution of 1%-3% when exposed between 100-600 MU min-1 (6 MV) and 3%-6% when exposed between 5-15 MV (600 MU min-1). While the stem effects were still observable, we expect that enhancing the current optical setup will simultaneously improve the scintillation signal and reduce the stem effects.

SUBMITTER: Kim TJ 

PROVIDER: S-EPMC7594590 | biostudies-literature | 2020 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Second window near-infrared dosimeter (NIR2D) system for radiation dosimetry.

Kim Tae Jin TJ   Cheng Kai K   Zhang Hao H   Liu Shi S   Skinner Lawrie L   Xing Lei L  

Physics in medicine and biology 20200831 17


Fiber-coupled scintillation dosimeters are a cost-effective alternative to the conventional ion chambers in radiation dosimetry. However, stem effects from optical fibers such as Cerenkov radiation incur significant errors in the readout signal. Here we introduce a second near-infrared window dosimeter, dubbed as NIR2D, that can potentially be used as real-time radiation detector for clinical megavoltage beams. Lanthanide-based rare-earth NaYF<sub>4</sub> nano-phosphors doped with both erbium an  ...[more]

Similar Datasets

| S-EPMC5679521 | biostudies-literature
| S-EPMC6137042 | biostudies-literature
| S-EPMC7696853 | biostudies-literature
| S-EPMC6907162 | biostudies-literature
| S-EPMC5622036 | biostudies-literature
| S-EPMC6506383 | biostudies-literature
| S-EPMC6348120 | biostudies-literature
| S-EPMC5293734 | biostudies-literature
| S-EPMC9145609 | biostudies-literature
| S-EPMC3358570 | biostudies-literature