Potential False-Positive and False-Negative Results for COVID-19 IgG/IgM Antibody Testing After Heat-Inactivation.


ABSTRACT: Objectives: With the worldwide spread of coronavirus disease 2019 (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), various antibody detection kits have been developed to test for SARS-CoV-2- specific IgG, IgM, and total antibody. However, the use of different testing methods under various heat-inactivation conditions might affect the COVID-19 detection results. Methods: Seven different antibody detection kits produced by four manufacturers for detection of SARS-CoV-2 IgG, IgM, and total antibody were tested at Wuhan Huoshenshan Hospital, China. Most of the kits used the indirect immunity, capture, and double-antigen sandwich methods. The effects of various heat-inactivation conditions on SARS-CoV-2-specific IgG, IgM, and total antibody detection were analyzed for the different test methods. Results: Using the indirect immunity method, values for SARS-CoV-2 IgG antibody significantly increased and those for IgM antibody decreased with increasing temperature of heat-inactivation using indirect immunity method. However, values for SARS-CoV-2 IgM and total antibody showed no change when the capture and double-antigen sandwich methods were used. The changes in IgG and IgM antibody values with the indirect immunity method indicated that heat-inactivation could affect COVID-19 detection results obtained using this method. In particular, 18 (22.2%) SARS-CoV-2 IgM positive samples were detected as negative with heat-inactivation at 65°C for 30 min, and one (25%) IgG negative sample was detected as positive after heat-inactivation at 56°C for 60 min and 60°C for 30 min. Conclusions: Heat-inactivation could increase SARS-CoV-2 IgG antibody values, and decrease IgM antibody values, causing potential false-positive or false-negative results for COVID-19 antibody detection using the indirect immunity method. Thus, before conducting antibody testing, the testing platforms should be evaluated in accordance with the relevant requirements to ensure accurate COVID-19 detection results.

SUBMITTER: Lin J 

PROVIDER: S-EPMC7849051 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Potential False-Positive and False-Negative Results for COVID-19 IgG/IgM Antibody Testing After Heat-Inactivation.

Lin Jie J   Dai Wei W   Li Weiwei W   Xiao Li L   Luo Tao T   Guo Yanju Y   Yang Yang Y   Han Ying Y   Zhu Peiran P   Wu Qiuyue Q   He Bangshun B   Wu Jian J   Xia Xinyi X  

Frontiers in medicine 20210118


<b>Objectives:</b> With the worldwide spread of coronavirus disease 2019 (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), various antibody detection kits have been developed to test for SARS-CoV-2- specific IgG, IgM, and total antibody. However, the use of different testing methods under various heat-inactivation conditions might affect the COVID-19 detection results. <b>Methods:</b> Seven different antibody detection kits produced by four manufacturers for dete  ...[more]

Similar Datasets

| S-EPMC8802144 | biostudies-literature
| S-EPMC7006501 | biostudies-literature
| S-EPMC3392728 | biostudies-other
| S-EPMC10722006 | biostudies-literature
| S-EPMC6078058 | biostudies-literature
| S-EPMC10430847 | biostudies-literature
| S-EPMC7794619 | biostudies-literature
| S-EPMC8257489 | biostudies-literature
| S-EPMC6783167 | biostudies-literature
| S-EPMC5501679 | biostudies-literature