Unknown

Dataset Information

0

Improving acetabular shell deformation testing.


ABSTRACT:

Background

To be FDA approved, acetabular shells must undergo rigorous testing. To prevent implant failure, acetabular shells must be able to tolerate peak loads during impaction with minimal deformation. The implants must therefore be validated in order to ensure that their structural integrity can tolerate peak loads. The current ISO 7206-12 recommends manufacturing an expensive single-purpose testing device to measure the deformation of acetabular shells. In the article, we provide an open access methodology, that can be conducted with minimal expense, for testing acetabular shell deformation.

Methods

We designed our experimental setup to utilize a servohydraulic materials testing device (Instron) commonly found in biomechanics laboratories and then validated the measurements optically with optical data. Furthermore, we designed an inexpensive acetabular screw fastener which acts as an adaptor that can be used to mount a variety of acetabular shell types just as effectively as current methods and is in compliance with the standards outlined in ISO 7206-12:2016.

Results

A Bland-Altman plot comparing the Instron and optical displacement measurements found the standard deviation of bias to be 0.046 mm and an insignificant systemic bias.

Conclusion

We have developed and validated a low-cost open-source system that can effectively test acetabular shell deformation that meets ISO standards.

SUBMITTER: Beitler BG 

PROVIDER: S-EPMC9246253 | biostudies-literature | 2022 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improving acetabular shell deformation testing.

Beitler Brian G BG   Crich Alexander A   Tommasini Steven M SM   Wiznia Daniel H DH  

Journal of the mechanical behavior of biomedical materials 20220401


<h4>Background</h4>To be FDA approved, acetabular shells must undergo rigorous testing. To prevent implant failure, acetabular shells must be able to tolerate peak loads during impaction with minimal deformation. The implants must therefore be validated in order to ensure that their structural integrity can tolerate peak loads. The current ISO 7206-12 recommends manufacturing an expensive single-purpose testing device to measure the deformation of acetabular shells. In the article, we provide an  ...[more]

Similar Datasets

| S-EPMC7286971 | biostudies-literature
| S-EPMC6811596 | biostudies-literature
| S-EPMC4957073 | biostudies-literature
| S-EPMC5585821 | biostudies-literature
| S-EPMC9867955 | biostudies-literature
| S-EPMC10903903 | biostudies-literature
| S-EPMC9576483 | biostudies-literature
| S-EPMC9386689 | biostudies-literature
| S-EPMC4958114 | biostudies-literature
| S-EPMC5503355 | biostudies-other