Standing surface acoustic wave (SSAW)-based microfluidic cytometer.


ABSTRACT: The development of microfluidic chip-based cytometers has become an important area due to their advantages of compact size and low cost. Herein, we demonstrate a sheathless microfluidic cytometer which integrates a standing surface acoustic wave (SSAW)-based microdevice capable of 3D particle/cell focusing with a laser-induced fluorescence (LIF) detection system. Using SSAW, our microfluidic cytometer was able to continuously focus microparticles/cells at the pressure node inside a microchannel. Flow cytometry was successfully demonstrated using this system with a coefficient of variation (CV) of less than 10% at a throughput of ~1000 events s(-1) when calibration beads were used. We also demonstrated that fluorescently labeled human promyelocytic leukemia cells (HL-60) could be effectively focused and detected with our SSAW-based system. This SSAW-based microfluidic cytometer did not require any sheath flows or complex structures, and it allowed for simple operation over a wide range of sample flow rates. Moreover, with the gentle, bio-compatible nature of low-power surface acoustic waves, this technique is expected to be able to preserve the integrity of cells and other bioparticles.

SUBMITTER: Chen Y 

PROVIDER: S-EPMC3956078 | biostudies-literature | 2014 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Standing surface acoustic wave (SSAW)-based microfluidic cytometer.

Chen Yuchao Y   Nawaz Ahmad Ahsan AA   Zhao Yanhui Y   Huang Po-Hsun PH   McCoy J Phillip JP   Levine Stewart J SJ   Wang Lin L   Huang Tony Jun TJ  

Lab on a chip 20140301 5


The development of microfluidic chip-based cytometers has become an important area due to their advantages of compact size and low cost. Herein, we demonstrate a sheathless microfluidic cytometer which integrates a standing surface acoustic wave (SSAW)-based microdevice capable of 3D particle/cell focusing with a laser-induced fluorescence (LIF) detection system. Using SSAW, our microfluidic cytometer was able to continuously focus microparticles/cells at the pressure node inside a microchannel.  ...[more]

Similar Datasets

| S-EPMC4442640 | biostudies-literature
| S-EPMC6291339 | biostudies-literature
| S-EPMC3997299 | biostudies-literature
| S-EPMC4188268 | biostudies-literature
| S-EPMC4688895 | biostudies-literature
| S-EPMC4856433 | biostudies-literature
| S-EPMC6423196 | biostudies-literature
| S-EPMC3365908 | biostudies-other
| S-EPMC9263997 | biostudies-literature
| S-EPMC8267599 | biostudies-literature