RF Safety Evaluation of a Breast Tissue Expander Device for MRI: Numerical Simulation and Experiment.


ABSTRACT: This study describes the MRI-related radio frequency (RF) safety evaluation of breast tissue expander devices to establish safety criteria. Numerical simulations and experimental measurements were performed at 64 MHz with a gel phantom containing a breast expander. Additionally, computational modeling was performed (64 and 128 MHz) with an adult female model, containing a virtually implanted breast tissue expander device for four imaging landmark positions. The presence of the breast tissue expander device led to significant alterations in specific absorption rate (SAR) and|B1+|distributions. The main source of SAR alterations with the use of the breast expander device was the saline-filled pouch of the expander. Conversely, the variation of RF magnetic field (B1+) was mainly caused by the metallic port. The measured values of electric field magnitude did not increase significantly due to the introduction of the expander device. The maximum 1g- or 10g-averaged SAR values in tissues near the implant were lower than those expected in other regions of the patient body with normalization of both|B1+|equal to 2 μT at the coil isocenter and whole body averaged SAR equal to 4W/kg.

SUBMITTER: Park BS 

PROVIDER: S-EPMC5814144 | biostudies-other | 2017 Oct

REPOSITORIES: biostudies-other

altmetric image

Publications

RF Safety Evaluation of a Breast Tissue Expander Device for MRI: Numerical Simulation and Experiment.

Park Bu S BS   Razjouyan Amir A   Angelone Leonardo M LM   McCright Brent B   Rajan Sunder S SS  

IEEE transactions on electromagnetic compatibility 20171001 5


This study describes the MRI-related radio frequency (RF) safety evaluation of breast tissue expander devices to establish safety criteria. Numerical simulations and experimental measurements were performed at 64 MHz with a gel phantom containing a breast expander. Additionally, computational modeling was performed (64 and 128 MHz) with an adult female model, containing a virtually implanted breast tissue expander device for four imaging landmark positions. The presence of the breast tissue expa  ...[more]

Similar Datasets

| S-EPMC6380517 | biostudies-literature
| S-EPMC7047541 | biostudies-literature
| S-EPMC5942199 | biostudies-literature
| S-EPMC6289759 | biostudies-literature
| S-EPMC7806143 | biostudies-literature
| S-EPMC7482415 | biostudies-literature
| S-EPMC6163945 | biostudies-literature
| S-EPMC7882116 | biostudies-literature
| S-EPMC6544887 | biostudies-literature
| S-EPMC8530882 | biostudies-literature